We are ready to share our knowledge of marketing worldwide and recommend you suitable products at most aggressive costs. So Profi Tools offer you finest benefit of money and we are ready to produce alongside one another with Competitive enzyme immunoassay Kit , QuinoSensor Test Kit , TetraSensor Test Kit , If you are interested in any of our products and services, please don't hesitate to contact us. We are ready to reply you within 24 hours after receipt of your request and to create mutual un-limited benefits and business in near future.
Factory directly C Reactive Protein Kit - HoneyGuard Tetracyclines Test Kit – kwinbon Detail:

Cat. KB01009K-50T

About
This kit is used for rapid qualitative analysis of tetracyclines in honey sample.

Sample prepare method
(1) If the honey sample crystallized ,warm it in water bath not high than 60℃, till the honey sample thaw ,mix completely ,cooling as room temperature , then weight for assay.
(2) Weigh 1.0±0.05g homogenate into a 10ml polystyrene centrifuge tube, add 3ml sample extraction solution, vortex for 2min or shake it by hand until the sample is mixed completely.

Assay operations.
(1.) Take bottles needed from the kit package, take out required cards, and make proper marks. Please use these test cards within 1h after open package.
(2.) Take 100l prepared sample into the sample hole by pipette, then start the timer after liquid flow.
(3.) Incubate for 10min at room temperature.
7.LOD

Tetracyclines

LOD(μg/L)

Tetracyclines

LOD(μg/L)

tetracycline

10

doxycycline 15
aureomycin

20

oxytetracycline

10

Results
There are 2 lines in the card result area, Control line and Tetracylcines Line, which are briefly  sed as “B” and “T”. The test results will depend on the color of these lines. The following diagram describes the result identification.
Negative: Control line and Test line both are red and T Line is darker than the control line;
Tetracyclines Positive: Control Line is red, T Line has no color or T Line is lighter color than C line, or the T Line is the same as C Line.

Results

Storage
2-30°C in dark dry place, do not freeze. The kit will be valid in 12 months. The lot number and expired date are printed on the package.


Product detail pictures:

Factory directly C Reactive Protein Kit - HoneyGuard Tetracyclines Test Kit – kwinbon detail pictures

Factory directly C Reactive Protein Kit - HoneyGuard Tetracyclines Test Kit – kwinbon detail pictures

Factory directly C Reactive Protein Kit - HoneyGuard Tetracyclines Test Kit – kwinbon detail pictures


Related Product Guide:

We've got a really efficient group to deal with inquiries from prospects. Our purpose is "100% customer fulfillment by our product excellent, price & our group service" and enjoy a superb track record amid clientele. With many factories, we can easily deliver a wide selection of Factory directly C Reactive Protein Kit - HoneyGuard Tetracyclines Test Kit – kwinbon , The product will supply to all over the world, such as: Zambia, Paraguay, Finland, Many kinds of different solutions are available for you to choose, you can do one-stop shopping here. And customized orders are acceptable. Real business is to get win-win situation, if possible, we would like to deliver more support for customers. Welcome all nice buyers communicate details of solutions with us!!
  • The company can think what our think, the urgency of urgency to act in the interests of our position, can be said this is a responsible company, we had a happy cooperation!
    5 Stars By King from London - 2018.05.22 12:13
    Adhering to the business principle of mutual benefits, we have a happy and successful transaction, we think we will be the best business partner.
    5 Stars By Christine from Jakarta - 2018.12.14 15:26
    Write your message here and send it to us